Professional Equipment and our suppliers are dedicated to providing quality products to our customers. We stand behind our commitment to quality with a satisfaction guarantee on every purchase*. If, for any reason, a customer is not satisfied with a product, they may return it within 30 days of the date purchased for a refund of the purchase price of the product(s). Returned merchandise must be in its original box, in new condition and include all instructions, and the invoice or packing slip.
*Exceptions: Videotapes, DVDs, audiotapes, computer software (once opened), animal traps (once opened), forms, reports, custom orders and leased items are not returnable. Books are returnable within 7 days of receipt. Test chemicals, once opened are not returnable. Downloads are non-refundable.
Confidently take reliable readings with the Fluke 353 True-RMS Clamp Meter. The extra-wide jaw easily clamps around large conductors, typically found in high-current applications. The extra rugged design and CAT IV 600 V, CAT III 1000 V ratings add an extra element of user protection when taking high-powered measurements. Accurate peak measurements can be taken using the in-rush current mode, ideal for motors and inductive loads. It also measures voltage and resistance, making this an ideal clamp meter for utilities, electrical contractors and industrial service technicians.
Features:
Reliably handle a wide range of high-current applications with 2000 A ac + dc true-rms, 1400 A ac, and 2000 A dc
The large 58 mm (2.3 in) jaw capacity is suitable for large or multiple conductors
CAT IV 600 V, CAT III 1000 V rating for added user protection
In-rush current measurement captures 'power-on’ surge current with accuracy and repeatability
Accurately measure frequency up to 1 kHz for optimum troubleshooting
Quickly analyze readings using the MIN, MAX, and AVG functions
A large backlit display allows for easy visibility in low-lit areas
Use the display hold feature to capture readings even when the display cannot be viewed
Use the low-pass filter to smooth out noisy loads and stabilize readings
Includes:
353 clamp meter
C43 Soft carrying case
6 AA batteries
User manual
3 year warranty
Spec Summary:
Electrical Specifications:
Current Measurement dc and ac 10 Hz to 100 Hz:
Range: 40 A
Resolution: 10 mA
Accuracy, A: 1.5 % rdg + 15 digits
Trigger Level for Inrush: 0.50 A
Trigger Level for Hz Filter OFF: 2.50 A
Trigger Level for Hz Filter ON: 0.50 A
Range: 400 A
Resolution: 100 mA
Accuracy, A: 1.5 % rdg + 5 digits
Trigger Level for Inrush: 5.0 A
Trigger Level for Hz Filter OFF: 2.5 A
Trigger Level for Hz Filter ON: 2.5 A
Range: 2000 A; 1400 ac rms
Resolution: 1 A
Accuracy, A: 1.5 % rdg + 5 digits
Trigger Level for Inrush: 5 A
Trigger Level for Hz Filter OFF: 8 A
Trigger Level for Hz Filter ON: 8 A
Crest Factor (50/60 Hz):
Range: 40 A
Crest Factor* : 2 @ 33 A, 2.4 @ 27 A
Range: 400 A
Crest Factor* : 2 @ 330 A, 2.4 @ 270 A
Range : 2000 A; 1400 ac rms
Crest Factor* : 2 @ 1000 A, 2.4 @ 833 A
Current Measurement ac 100.1 Hz to 1 kHz:
Range: 40 A
Resolution: 10 mA
Accuracy > 10 A : 3.5 % rdg + 15 digits
Trigger Level for Inrush: 0.50 A
Trigger Level for Hz Filter OFF : 2.50 A
Trigger Level for Hz Filter ON : 0.50 A
Range: 400 A
Resolution: 100 mA
Accuracy > 10 A : 3.5 % rdg + 5 digits
Trigger Level for Inrush: 5.0 A
Trigger Level for Hz Filter OFF : 2.5 A
Trigger Level for Hz Filter ON : 2.5 A
Range: 2000 A; 1400 ac rms
Resolution: 1 A
Accuracy > 10 A : 3.5 % rdg + 5 digits
Trigger Level for Inrush: 5 A
Trigger Level for Hz Filter OFF : 8 A
Trigger Level for Hz Filter ON : 8 A
Frequency Measurement:
Measurement Range Resolution:
5.0 Hz to 1 kHz
0.1 Hz (15 Hz to 399.9 Hz); 1 Hz (400 Hz to 1 kHz)
Accuracy – 5.0 Hz to 100 Hz:
0.2 % + 2 counts
Accuracy – 100.1 Hz to 1 kHz:
0.5 % + 5 counts
Trigger level:
Refer to current and voltage tables
*Add 2 % to error spec for CF > 2
General Specifications:
Batteries:
Six 1.5 V AA NEDA 15 A or IEC LR6
Battery life (with typical usage, backlight off):
100 hours
Test leads:
Rated to 1000 V
Weight:
.814 kg (1.8 lb)
Jaw size:
58 mm (2.28 in)
Dimensions (LxWxD):
300 mm x 98 mm x 52 mm (12 in x 3.75 in x 2 in)
Safety rating:
IEC 61010-2-032, 600 V CAT IV, 1000 V CAT III
Environmental Specifications:
Operating Temperature:
32 °F to + 122 °F (0 °C to +50 °C)
Storage Temperature:
-4 °F to 140 °F (-20 °C to +60 °C)
Operating Humidity:
0 to 95% (non-condensing)
Operating Altitude:
2000 m
Storage Altitude:
10,000 m
IP Rating:
42 (indoor use only)
Drop Test Requirements:
1 m
EMI, RFI, EMC:
FCC part 15, IEC/EN 61326-1:1997 class B, IEC/EN 61326:1997 3V/m, performance criteria B, EN61325