Professional Equipment and our suppliers are dedicated to providing quality products to our customers. We stand behind our commitment to quality with a satisfaction guarantee on every purchase*. If, for any reason, a customer is not satisfied with a product, they may return it within 30 days of the date purchased for a refund of the purchase price of the product(s). Returned merchandise must be in its original box, in new condition and include all instructions, and the invoice or packing slip.
*Exceptions: Videotapes, DVDs, audiotapes, computer software (once opened), animal traps (once opened), forms, reports, custom orders and leased items are not returnable. Books are returnable within 7 days of receipt. Test chemicals, once opened are not returnable. Downloads are non-refundable.
The Fluke 707 Loop Calibrator is a high performance solution for calibration, repair and maintenance of current loops. The new "Quick Click" knob making it extremely fast and easy to use. It features:
Features:
Large display and simple, quick-click interface for easy one-handed operation
Simultaneous mA and % readout for quick, easy interpretation of readings
mA accuracy of 0.015%, superior to other loop calibrators in this price range
1 µA resolution for mA source, simulate and measure
Push button 25% steps for fast, easy linearity checks
0-100% span check for fast confirmation of zero and span
Selectable slow ramp, fast ramp, and step ramp to provide smooth outputs for valve slewing and loop functional tests
24 volt internal loop supply, so you can power and read a transmitter at the same time without carrying a DMM
Measures V dc to 28 V
Single 9V battery that is easily changed without braking calibration seal
0 - 20 mA or 4 - 20 mA default start up modes
Innovative output adjustment dial with 1 µA and 100 resolution
HART mode connects 250 Ù resistor in series with 24 V loop for compatibility with HART communicators
Spec Summary:
Specifications
Current Measurement:
Range: 0 to 24 mA
Resolution: 0.001 mA
Accuracy: 0.015% Rdg + 2 LSD
Current Sourcing:
Range: 0 to 20 mA or 4 to 20 mA
Accuracy: 0.015% Rdg + 2 LSD
Drive Capability: 1200 Ù; 950 Ù at 20 mA in HART mode
Current Simulation:
Range: 0 to 20 mA or 4 to 20 mA
Accuracy: 0.015% Rdg + 2 LSD
Ext. loop voltage requirement: 12-30 V
Loop Power:
While measuring mA, 24 V
Voltage measurement:
Range: 0 to 28 V DC
Resolution: 1 mV
Accuracy: 0.015% Rdg + 2 LSD
Environmental Specifications
Operating Temperature:
14 to 131°F (-10 to 55°C)
Storage Temperature:
-22 to 140°F (-30 to 60°C)
Humidity (Without Condensation):
0 to 90% 32 to 95°F (0 to 35°C)
0 to 70% 95 to 131°F (35 to 55°C)
Temperature Coefficient:
14 to 64.4°F (-10 to 18°C),
82.4 to 131°F (28 to 55°C, )
±.005% of range per °C
Safety Specifications
Safety:
CSA C22.2 No. 1010.1:1992
EMC:
EN50082-1:1992 and EN55022:1994 Class B
Maximum voltage:
30V (protection circuit for 230V)
Mechanical & General Specifications
Size:
5.98" x 2.75" x 1.61" (152 x 70 x 41 mm)
Weight:
7.9 oz (11.9 oz) [224 g (340 g including holster)]