Hanna Waterproof Multiparameter Water Quality Meter - High Range
This Hanna waterproof multiparameter water quality meter has a housing that is rated for IP 67 conditions
Item #: 57566
Manufacturer: Hanna Instruments Incorporated
Manufacturer Item #: HI991301
Your Price $509.95
Features & Applications:
The Hanna Waterproof Multiparameter Water Quality Meter is an easy-to-use instrument that tests pH, conductivity (EC), Total Dissolved Solids (TDS) and temperature with a single probe. The waterproof Water Quality Meter utilizes gel electrolyte and a cloth junction for the pH sensor, and graphite for the EC/TDS sensor that eliminates oxidation and polarization errors. Features include auto-calibration and auto-temperature compensation, and the two buttons control all the functions, including the selection of calibration buffers and the temperature scale (°C or °F).The unit also includes different TDS factors from 0.45 to 1.00, and a range of temperature coefficients (ß) from 0.0 to 2.4% for greater consistency and reproducibility. Users also have the ability to select standardized buffer recognition values. Additionally, a solid-state amplifier is integrated into the probe to ensure against interference from transient electrical noise. The Hanna Waterproof Multiparameter Water Quality Meter is rated for IP 67 conditions and includes a probe and a hard carrying case.
Specifications:
| Range: | pH | 0.00 to 14.00 pH |
| Range: | EC | 0.00 to 20.00 mS/cm |
| Range: | TDS | 0.00 to 10.00 ppt (g/L) |
| Range: | Temperature | 0.0 to 60.0°C (32.0 to 140.0°F) |
| Resolution: | pH | 0.01 pH |
| Resolution: | EC | 0.01 mS/cm |
| Resolution: | TDS | 0.01 ppt |
| Resolution: | Temperature | 0.1°C (0.1°F) |
| Accuracy: | pH | ±0.01 pH |
| Accuracy: | EC | ±2% F.S. |
| Accuracy: | TDS | ±2% F.S. |
| Accuracy: | Temperature | ±0.5°C (±1.0°F) |
| TDS Conversion Factor: | selectable 0.45 to 1.00 with 0.01 increments (default 0.50) | |
| pH Calibration: | automatic at 1 or 2 points with 2 sets of memorized standard (pH 4.01/7.01/10.01 or 4.01/6.86/9.18) | |
| EC/TDS Calibration: | automatic, 1 point at 6.44 ppt (CONV 0.5) at 9.02 ppt (CONV 0.7) at 12880 µS/cm (others) |
|
| Temperature Compensation: | pH | Automatic |
| Temperature Compensation: | EC/TDS | Automatic with selectable ß from 0.0 to 2.4%/°C with 0.1 increments |
| Probe: | HI 1288, pH/EC/TDS/T, DIN connector, 1 m (3.3') cable (included) | |
| Battery Type / Life: | 4 x 1.5V AAA / approx. 500 hours of continuous use | |
| Environment: | 0 to 50°C (32 to 122°F); RH max 100% | |
| Dimensions: | 150 x 80 x 36 mm (5.9 x 3.1 x 1.4") | |
| Weight: | 210 g (7.4 oz.) |
| Product Description: | Price: | Your Cart: | |
|---|---|---|---|
| Replacement Probe | $124.95 | Add | |