| Infrared Temperature Range: |
-40 °C to 800 °C (-40 °F to 1472 °F) |
| Infrared Accuracy: |
< 0 °C (32 °F): ± (1.0 °C (± 2.0 °F) + 0.1°/1 °C or °F);
> 0 °C (32 °F): ± 1 % or ± 1.0 °C (± 2.0 °F), whichever is greater |
| Display Resolution: |
0.1 °C / 0.1 °F |
| Infrared Spectral Response: |
8 µm to 14 µm |
| Infrared Response Time: |
< 500 msec |
| Thermocouple Type-K Input Temperature Range: |
-270 °C to 1372 °C (-454 °F to 2501 °F) |
| Thermocouple Type-K Input Accuracy: |
-270 °C to -40 °C: ± (1 °C + 0.2 °/1 °C) (-454 °F to -40 °F: ± (2 °F + 0.2 °/1 °F
-40 °C to 1372 °C: ± 1 % or 1 °C (-40 °F to 2501 °F: ± 1 % or 2 °F), whichever is greater |
| D:S (Distance to measurement spot size: |
50:1 |
| Laser Sighting: |
Single-point laser < 1 mw output Class 2 (II) operation, 630 nm to 670 nm |
| Minimum Spot Size: |
19 mm (0.75 in) |
| Emissivity Adjustment: |
By built-in table of common materials or digitally adjustable from 0.10 to 1.00 by 0.01 |
| Data storage w/ Date/Time Stamp: |
99 points |
| PC Interface and Cable: |
USB 2.0 with FlukeView Forms software |
| Hi/Low Alarms: |
Audible and two-color visual |
| Min/Max/Avg/Dif: |
Yes |
| Display: |
Dot matrix 98 x 96 pixels with functions menus |
| Backlight: |
Two levels, normal and extra bright for darker environments |
| Trigger Lock: |
Yes |
| Switchable Celsius and Fahrenheit: |
Yes |
| Power: |
2AA/LR6 batteries and USB when used with a PC |
| Battery Life: |
If used continuously: laser and backlight on, 12 hours; laser and backlight off, 100 hours |
| Operating Temperature: |
0 °C to 50 °C (32 °F to 122 °F) |
| Storage Temperature: |
-20 °C to 60 °C (-40 °F to 140 °F) |
| Bead Thermocouple Type-K Range: |
-40 °C to 260 °C (-40 °F to 500 °F) |
| Bead Thermocouple Type-K Accuracy: |
± 1.1 °C (2.0 °F) from 0 °C to 260 °C (32 °F to 500 °F), typically within 1.1 °C (2.0 °F) from -40 °C to 0 °C (-40 °F to 32 °F) |