The Fluke 353 True-RMS Clamp Meter features a large backlit display for easy visibility in low-lit areas!
Features & Applications:
Confidently take reliable readings with the Fluke 353 True-RMS Clamp Meter. The extra-wide jaw easily clamps around large conductors, typically found in high-current applications. The extra rugged design and CAT IV 600 V, CAT III 1000 V ratings add an extra element of user protection when taking high-powered measurements. Accurate peak measurements can be taken using the in-rush current mode, ideal for motors and inductive loads. It also measures voltage and resistance, making this an ideal clamp meter for utilities, electrical contractors and industrial service technicians.
Features:
- Reliably handle a wide range of high-current applications with 2000 A ac + dc true-rms, 1400 A ac, and 2000 A dc
- The large 58 mm (2.3 in) jaw capacity is suitable for large or multiple conductors
- CAT IV 600 V, CAT III 1000 V rating for added user protection
- In-rush current measurement captures 'power-on’ surge current with accuracy and repeatability
- Accurately measure frequency up to 1 kHz for optimum troubleshooting
- Quickly analyze readings using the MIN, MAX, and AVG functions
- A large backlit display allows for easy visibility in low-lit areas
- Use the display hold feature to capture readings even when the display cannot be viewed
- Use the low-pass filter to smooth out noisy loads and stabilize readings
- Includes:
- 353 clamp meter
- C43 Soft carrying case
- 6 AA batteries
- User manual
- 3 year warranty
Spec Summary:
| Electrical Specifications: |
| Current Measurement dc and ac 10 Hz to 100 Hz: |
Range: 40 A
|
Resolution: 10 mA
Accuracy, A: 1.5 % rdg + 15 digits
Trigger Level for Inrush: 0.50 A
Trigger Level for Hz Filter OFF: 2.50 A
Trigger Level for Hz Filter ON: 0.50 A |
| Range: 400 A |
Resolution: 100 mA
Accuracy, A: 1.5 % rdg + 5 digits
Trigger Level for Inrush: 5.0 A
Trigger Level for Hz Filter OFF: 2.5 A
Trigger Level for Hz Filter ON: 2.5 A |
Range: 2000 A; 1400 ac rms
|
Resolution: 1 A
Accuracy, A: 1.5 % rdg + 5 digits
Trigger Level for Inrush: 5 A
Trigger Level for Hz Filter OFF: 8 A
Trigger Level for Hz Filter ON: 8 A
|
| Crest Factor (50/60 Hz): |
Range: 40 A
|
Crest Factor* : 2 @ 33 A, 2.4 @ 27 A |
| Range: 400 A |
Crest Factor* : 2 @ 330 A, 2.4 @ 270 A |
| Range : 2000 A; 1400 ac rms |
Crest Factor* : 2 @ 1000 A, 2.4 @ 833 A |
| Current Measurement ac 100.1 Hz to 1 kHz: |
Range: 40 A |
Resolution: 10 mA
Accuracy > 10 A : 3.5 % rdg + 15 digits
Trigger Level for Inrush: 0.50 A
Trigger Level for Hz Filter OFF : 2.50 A
Trigger Level for Hz Filter ON : 0.50 A |
| Range: 400 A |
Resolution: 100 mA
Accuracy > 10 A : 3.5 % rdg + 5 digits
Trigger Level for Inrush: 5.0 A
Trigger Level for Hz Filter OFF : 2.5 A
Trigger Level for Hz Filter ON : 2.5 A |
| Range: 2000 A; 1400 ac rms |
Resolution: 1 A
Accuracy > 10 A : 3.5 % rdg + 5 digits
Trigger Level for Inrush: 5 A
Trigger Level for Hz Filter OFF : 8 A
Trigger Level for Hz Filter ON : 8 A |
| Frequency Measurement: |
Measurement Range Resolution: |
5.0 Hz to 1 kHz
0.1 Hz (15 Hz to 399.9 Hz); 1 Hz (400 Hz to 1 kHz) |
| Accuracy – 5.0 Hz to 100 Hz: |
0.2 % + 2 counts |
| Accuracy – 100.1 Hz to 1 kHz: |
0.5 % + 5 counts |
| Trigger level: |
Refer to current and voltage tables |
*Add 2 % to error spec for CF > 2
| General Specifications: |
| Batteries: |
Six 1.5 V AA NEDA 15 A or IEC LR6 |
| Battery life (with typical usage, backlight off): |
100 hours |
| Test leads: |
Rated to 1000 V |
| Weight: |
.814 kg (1.8 lb) |
| Jaw size: |
58 mm (2.28 in) |
| Dimensions (LxWxD): |
300 mm x 98 mm x 52 mm (12 in x 3.75 in x 2 in) |
| Safety rating: |
IEC 61010-2-032, 600 V CAT IV, 1000 V CAT III |
| Environmental Specifications: |
| Operating Temperature: |
32 °F to + 122 °F (0 °C to +50 °C) |
| Storage Temperature: |
-4 °F to 140 °F (-20 °C to +60 °C) |
| Operating Humidity: |
0 to 95% (non-condensing) |
| Operating Altitude: |
2000 m |
| Storage Altitude: |
10,000 m |
| IP Rating: |
42 (indoor use only) |
| Drop Test Requirements: |
1 m |
| EMI, RFI, EMC: |
FCC part 15, IEC/EN 61326-1:1997 class B, IEC/EN 61326:1997 3V/m, performance criteria B, EN61325 |
| Temperature Coefficients: |
Current:
|
0.1 % of reading per °C outside 22 °C to 24 °C |
| Voltage: |
0.1 % of reading per °C outside 22 °C to 24 °C |
|