Equipped with Telaire's patented dual beam Absorption Infrared technology, the easy-to-use Telaire Handheld Indoor Air Quality Monitor is designed for use in residential or commercial applications. The 7001 has the ability to display CO2 readings and calculate outside air ventilation rates in cfm/person. It is the ideal tool for identifying energy saving opportunities in over-ventilated spaces, determining if air quality complaints are due to insufficient ventilation, or locating the presence of combustion fumes generated from vehicles and appliances. A power adapter is included.
| Method: |
Dual Beam Absorption Infrared |
| Sample Method: |
Diffusion or flow through (50 to 100 ml/min) |
| Display-LCD: |
Independent CO2 and temperature readings. Calculates and displays ventilation rates. |
| General: |
| Operating Conditions: |
32°F to 122°F (0°C to 50°C)
0 to 95% RH, non-condensing |
| Storage Temperatures: |
-4°F to 140°F (-20°C tp 60°C) |
| Certification: |
FCC Class 15 Part B |
| Warranty: |
18 months parts and labor |
| Performance-CO2 Channel: |
| Measurement Range: |
0 to 10,000 ppm display
0 to 4000 ppm voltage output |
| Display Resolution: |
±1 ppm |
| Accuracy: |
±50 ppm or 5% of reading, whichever is greater |
| Repeatability: |
±20 ppm |
Temperature Dependence:
|
±0.1% of reading per °C or ±2 ppm per °C, whichever is greater, referenced at 25°C. |
| Pressure Dependence: |
0.13% of reading per mmHg (corrected via user input for
elevation) |
| Response Time: |
<60 seconds for 90% of step change |
| Warm-Up Time: |
<60 seconds at 72°F (22°C) |
| Calibration Interval: |
12 months, offset adjustment using single gas at 0 to 1000 ppm CO2. Full factory calibration available. |
| Performance-Temperature Channel: |
| Temperature Range: |
Voltage output: 32°F to 104°F (0°C to 40°C)
Display: 32°F to 122°F (0°C to 50°C) |
| Display Resolution: |
0.1°F (0.1°C) |
| Display Options: |
°F, °C, of off. Set with panel button. |
| Accuracy: |
±2°F (±1°C) |
| Response Time: |
20 to 30 minutes (case must equilibrate with
environment) |
| Calibration Interval: |
12 months, offset adjustment using temperature standard at 50°F to 86°F (10°C to 30°C). Full factory calibration available. |
| Output-Analog: |
| CO2: |
0 to 4 VDC, 1 mV/ppm (4000 ppm maximum) |
| Temperature: |
0 to 4 VDC linear, 32°F to 104°F (0°C to 40°C) |
| Output Impedance: |
100 ohms |
| Digital Output: |
RS232 for use with Telaire CO2View graphing software |
| Wiring Connection: |
One RJ45 to connector dual analog output plus digital
output |
| Power Supply: |
| Battery Type: |
Four AA batteries (not included) |
| Battery Operation: |
80 hours (alkaline) |
| External: |
6 VDC from external AC/DC adapter (included) |
| Power Requirements: |
100 mA peak, 20 mA average from 6V |