Easily save your surveys for analysis and sharing with this Wireless Handheld Radiation Alert Detector!
Features & Applications:
This Wireless Handheld Radiation Alert Detector allows you to complete your surveys automatically with your Abacus and a Windows Mobile, Palm Pilot or Windows Computer. Not only does it offer excellent sensitivity to low levels of alpha, gamma, and x-rays, but the internal "guard" detector automatically performs gamma subtraction from the "pancake" detector for unprecedented beta sensitivity. The software can display a variety of units for a wide array of applications, take timed counts, control the audio of each tube, and save your surveys for easy analysis and sharing, all wirelessly with Bluetooth right to your computer! It also has a red count light, a beeper sound and selectable alarm thresholds for quick assessments.
*REQUIRES A PC / SMART PHONE / PALM PILOT / LAPTOP THAT HAS WINDOWS XP / 98 / 2000 AND HAS BLUETOOTH CAPABILITES IN ORDER TO DISPLAY READINGS.
Features:
- Excellent sensitivity to low levels of alpha, gamma, and x-rays
- Internal "guard" detector automatically performs gamma subtraction from the "pancake" detector
- Displays a variety of units
- Timed counts
- Ability to control the audio of each tube
- Ability to save your surveys for easy analysis and sharing
- Red count light
- Beeper sound
- Selectable alarm thresholds
Spec Summary:
| Detector: |
Internal Tube 1: Halogen-quenched, uncompensated GM pancake tube with thin mica window, 1.4-2.0 mg/cm2 areal density. Effective diameter of window is 45 mm (1.75 in.) |
Internal Tube 2: Halogen-quenched, uncompensated thin wall beta-gamma GM peanut tube Radiation symbol on front and top labels mark the center of the detectors |
| Display: |
Windows Mobile PocketPCs, Palm Pilots or Windows based laptop or desktop computer |
| Operating Range Exposure Rate: |
µR/h - 0 to 2,000,000 µR/h
mR/h - 0 to 2,000 mR/h
µSv/h - 0 to 20,000 µSv/h |
| Count Rate: |
cpm - 0 to 750,000 cpm
cps - 0 to 12,500 cps |
| Sample Counting: |
Total counts - >99,999,999,999 |
Accumulated Dose:
|
1.8E308 |
| Total Time: |
Unlimited |
| Accuracy: |
Typically ±15% of reading |
| Alert Set Range: |
mR/hr .001 - 50 and CPM 1 - 160,000 |
| Gamma Sensitivity: |
7.5 CPS/mR/hr referenced to Co60. Smallest detectable level for I-125 is .02 µCi at contact |
| Timer: |
Total Count timer from 1 -99,999,999,999 seconds |
| Count Light: |
Red LED and internal beeper (can be switched off for silent operation) |
| Outputs: |
Bluetooth Wireless Serial Connection |
| Anti-Saturation: |
Meter will hold at full scale in fields as high as 100 times the maximum reading |
| Power Requirements: |
Two (2) AA Batteries |
| Temperature Range: |
-10ºC to 50ºC (14ºF to 122ºF) |
| Weight: |
337.4 grams (11.9 oz.) with battery |
| Size: |
332.7 x 63.5 x 35.6 mm (13.1 x 2.5 x 1.4in.) |
| Warranty: |
90 day Warranty on GM Detector, 1 year limited |
|