This Hanna multiparameter GPS enabled water quality meter has an amazing range of useful features that simply cannot be beat. Along with quick calibration, a built-in barometer, measurement check, automatic recognition of pH/ORP probe and USB connectivity, the meter can also perform on-demand or automatic logging of up to 60,000 samples with 13 measurements each. Designed for outdoor environments, the meter is impact resistant and waterproof meeting IP67 standards (30 minutes immersion under 1 m of water). The multi sensor probe can be left underwater (IP68 standard).
GPS information can be transferred to a PC using HANNA’s HI 929828 software and viewed by GPS mapping software. The built-in 12 channel GPS receiver and antenna guarantees a position accuracy of 30 ft (10 m). For measuring points within a 30 ft range or where GPS signals are not available, HI 9828’s Fast Tracker is invaluable for associating measurements with their locations.
HANNA’s exclusive Fast Tracker-Tag I.D. System monitors and records data using iButton®s that can be installed at any number of sampling sites. The HI 9828 has a graphic, backlit display that automatically sizes the digits to fit the screen and allows full configuration of each parameter measured along with on-screen graphing capability. The meter incorporates comprehensive GLP features and the downloading of data via USB connection. Each parameter is supported by on-screen context sensitive help, both in the calibration mode and during measurement.
The new HI 9828 features an internal 12 channel GPS receiver and antenna that calculates its position to track locations along with measurement data. The GPS tracks your location using satellites to within 30 ft (10 m) so you can be sure that you return to the same location for repeated measurements.
Users can connect to GPS tracking software such as Google Maps to view locations where samples have been taken. Measurement information is shown right on the map.
HANNA’s iButton® Tag Identification System simplifies test logging. iButton®s with a unique ID can be installed at various sampling sites. When the matching connector on the meter contacts the location button, measurements are logged and labeled with the alphanumeric user-entered location ID. Location, date, time and measurements are logged into the meter which can be transferred to a PC. The Fast Tracker system complements the GPS for ultimate tracking.
| Range: |
pH |
0.00 to 14.00 pH |
| Resolution: |
pH |
0.01 pH |
| Accuracy: |
pH |
±0.02 pH |
| Range: |
mV of pH input |
±600.0 mV |
| Resolution: |
mV of pH input |
0.1 mV |
| Accuracy: |
mV of pH input |
±0.5 mV |
| Range: |
ORP |
±2000.0 mV |
| Resolution: |
ORP |
0.1 mV |
| Accuracy: |
ORP |
±1.0 mV |
| Range: |
Dissolved Oxygen |
0.0 to 500.0% / 0.00 to 50.00 mg/L |
| Resolution: |
Dissolved Oxygen |
0.1% / 0.01 mg/L |
| Accuracy: |
Dissolved Oxygen |
0.0 to 300.0%: ±1.5% of reading or ±1.0% whichever is greater; 300.0 to 500.0%: ±3% of reading; 0.00 to 30.00 mg/L: ±1.5% of reading or 0.10 mg/L whichever is greater; 30.00 mg/L to 50.00 mg/L: ±3% of reading |
| Range: |
Conductivity |
0.000 to 200.000 mS/cm (absolute EC up to 400 mS/cm) |
| Resolution: |
Conductivity |
Manual: 1 µS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm; 1 mS/cm; Automatic: 1 µS/cm from 0 to 9999 µS/cm; 0.01 mS/cm from 10.00 to 99.99 mS/cm ; 0.1 mS/cm from 100.0 to 400.0 mS/cm; Automatic mS/cm: 0.001 mS/cm from 0.000 to 9.999 mS/cm; 0.01 mS/cm from 10.00 to 99.99 mS/cm; 0.1 mS/cm from 100.0 to 400.0 mS/cm |
| Accuracy: |
Conductivity |
±1% of reading or ±1 µS/cm whichever is greater |
| Range: |
Resistivity |
0 to 999999 Ωcm; 0 to 1000.0 kΩcm; 0 to 1.0000 MΩcm |
| Resolution: |
Resistivity |
Dependent on resistivity reading |
| Range: |
TDS |
0 to 400000 mg/L or ppm (the maximum value depends on the TDS factor) |
| Resolution: |
TDS |
Manual: 1 mg/L (ppm); 0.001 g/L (ppt); 0.01g/L (ppt); 0.1 g/L (ppt); 1 g/L (ppt); Auto-range scales: 1 mg/L (ppm) from 0 to 9999 mg/L (ppm); 0.01 g/L (ppt) from 10.00 to 99.99 g/L (ppt); 0.1 g/L (ppt) from 100.0 to 400.0 g/L (ppt); Auto-range g/L (ppt) scales: 0.001 g/L (ppt) from 0.000 to 9.999 g/L (ppt); 0.01 g/L (ppt) from 10.00 to 99.99 g/L (ppt); 0.1 g/L (ppt) from 100.0 to 400.0 g/L (ppt) |
| Accuracy: |
TDS |
±1% of reading or ±1 mg/L |
| Range: |
Salinity |
0.00 to 70.00 PSU (Extended Practical Salinity Scale) |
| Resolution: |
Salinity |
0.01 PSU |
| Accuracy: |
Salinity |
±2% of reading or 0.01 PSU whichever is greater |
| Range: |
Seawater Specific Gravity |
0.0 to 50.0 σt, σ0, σ15 |
| Resolution: |
Seawater Specific Gravity |
0.1 σt, σ0, σ15 |
| Accuracy: |
Seawater Specific Gravity |
±1 σt, σ0, σ15 |
| Range: |
Atm. Pressure |
450 to 850 mmHg; 17.72 to 33.46 inHg; 600.0 to 1133.2 mbar; 8.702 to 16.436 psi; 0.5921 to 1.1184 atm; 60.00 to 113.32 kPa |
| Resolution: |
Atm. Pressure |
0.1 mmHg; 0.01 inHg; 0.1 mbar; 0.001 psi; 0.0001 atm; 0.01 kPa |
| Accuracy: |
Atm. Pressure |
±3 mmHg within ±15°C from the temperature during calibration |
| Range: |
Temperature |
-5.00 to 55.00°C; 23.00 to 131.00°F; 268.15 to 328.15K |
| Resolution: |
Temperature |
0.01°C; 0.01°F; 0.01K |
| Accuracy: |
Temperature |
±0.15°C; ±0.27°F; ±0.15K |
| Calibration: |
pH |
Automatic 1, 2, or 3 points with 5 memorized standard buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01) or 1 custom buffer |
| Calibration: |
ORP |
Automatic at 1 custom point |
| Calibration: |
Conductivity, Salinity |
Automatic 1 point with 6 memorized standards (84 μS/cm, 1413 μS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm) or custom point |
| Calibration: |
Dissolved Oxygen |
Automatic 1 or 2 points at 0, 100% or 1 custom point |
| Calibration: |
Resistivity, TDS |
Based on conductivity or salinity calibration |
| Calibration: |
Atm. Pressure |
Automatic at 1 custom point |
| Calibration: |
Temperature |
Automatic at 1 custom point |
| Temperature Compensation: |
|
Automatic from -5 to 55°C (23 to 131°F) |
| Logging Memory: |
|
Up to 60000 samples with 13 measurements each Up to 45,000 samples with 15 measurements each (GPS) |
| Logging Interval: |
|
1 second to 3 hours |
| Computer Interface: |
|
USB (with HI 929828 software) |
| Waterproof Protection: |
|
Meter IP67, Probe IP68 |
| Environment: |
|
0 to 50°C (32 to 122°F); RH 100% |
| Power Supply: |
|
(4) 1.5V alkaline C cells (approx. 150 hours of continuous use without backlight) /(4) 1.2V rechargeable C cells (approx. 70 hours of continuous use without backlight) |
| Dimensions Meter: |
|
221 x 115 x 55 mm (8.7 x 4.5 x 2.2) |
| Dimensions Probe: |
|
270 x 46 mm DIA (10.6 x 1.8"DIA) |
| Weight: |
|
Meter: 750g (26.5 oz.); Probe: 750g (26.5 oz.) |