This Fieldpiece diagnostic digital sling psychrometer has a rubber boot helps protect the meter during bangs and drops
Features & Applications:
This Fieldpiece diagnostic digital sling psychrometer measures temperature, relative humidity, wet bulb and dew point. Additionally, the SRH2 calculates target superheat (TSH) and target evaporator exit temperature (TEET). The unit features easy calibration capabilities so that you can perform quick calibration in the field, and the built-in TEET alerts you to airflow problems across the evaporator. A superheat analysis allows the technician to determine if the refrigerant charge is correct on a system with a fixed restrictor.
Features
- Wet bulb,dry bulb, dew point and %RH
- Target Superheat and Target Evaporator Exit Temperature
- For temperature measurements, select either the built-in thermistor or plug-in thermocouple (included)
- Max/Min/Hold functions
- Blue backlight
- Rubber boot helps protect the meter during bangs and drops
- Magnetic hanger for hands free operation
- Standard 9V battery
- Auto-off to save battery life
- Thermocouple can be easily calibrated in the field
Benefits:
- It's a digital psychrometer.
- It's a diagnostic tool with TSH and TEET built in. No charts to fumble with.
- Choose the NTC or the included WB and DB T/Cs for testing, You decide.
- Faster testing and system diagnosis.
- Higher quality job.
- Test multiple parameters with one instrument.
- All this for about the same cost as competitive digital psychrometers.
Specifications:
| Operating environment: |
0°F - 122°F at <75%RH. |
| Temperature (K-type Thermocouple): |
| Range: |
58°F to 1832°F (-50°C to 1000°C) |
| Accuracy: |
±(0.1% rdg+2°F) -58°F to 1832°F
±(0.1% rdg+1°C) -50°C to 1000°C |
| Temperature (NTC): |
| Range: |
32°F to 140°F (0°C to 60°C) |
| Accuracy: |
±1°F from 50°F to 113°F (±0.5°C 10°C to 45°C)
±4°F from 32°F to 50°F and 113°F to 140°
±2°C from 0°C to 10°C and 45°C to 60°C |
| Relative Humidity: |
| Range: |
0% to 100% RH. |
| Accuracy: |
±2.5% at 77°F (25°C), 10% to 90%RH
±5% at 77°F (25°C) 0 to 10% RN & 90 to 100%RH |
| Battery: |
9V (included). |
| Battery life: |
Approx. 150 hrs (alkaline). |
| Auto-off: |
Automatically turns off after 15 minutes. (Can be disabled) |
| Includes: |
Meter, (ATA1) k-type dry bulb thermocouple w/clip, (ATWB1) k-type wet bulb thermocouple
w/clip, battery and Operators Manual. |
|