Professional Equipment     Print Page
PO Box 5197
Janesville, WI 53547-5197

Phone Number: 608-743-8002
Fax Number: 608-743-8039
Toll Free Number: 1-800-334-9291
Toll Free Fax Number: 1-888-776-3187


Kanomax Climomaster Series Hot Wire Anemometer

Kanomax Climomaster Series Hot Wire Anemometer
Item #: HV508-0531
Manufacturer: Kanomax
Manufacturer Item #: A531
Product Type: Anemometers
Your Price: $1,594.95
Kanomax Climomaster Series Hot Wire Anemometer stores up to 1500 measurement data

Features & Applications:

The world's most accurate handheld hot-wire anemometer in it's class.

Features:

  • Probe compatability feature allows you to have spare probe
  • Robust design
  • Detachable probe
  • Automatic Flow Rate Calculation function
  • RS232C terminal for data logging

Applications:

  • Ventilation testing
  • Laboratory control
  • Cleanrooms
  • IAQ investigation
  • Industrial hygiene
  • Quality control

Spec Summary:

Probe Type: Bi-directional (Bar type)
Measuring object: Airflow at ambient temperature and humidity
Air Velocity: Measuring Range: 20 to 6,000fpm
Resolution: 1fpm
Accuracy: ±2% of reading or ±3fpm whichever is greater
Response Time: Approximately 1 second
Temperature: Measuring Range: 2.0 to 140.0°F
Resolution: 0.1°F
Accuracy: ±1°F
Response Time: Approximately 30 seconds
Relative Humidity: Resolution: 0.1%RH
Accuracy: ±2.0%RH from 2 to 80%RH
±3%RH from 80 to 98%RH
Response Time: Approximately 15 seconds
Functions: Max., Min., Average, Hold, max Hold, Time Constant (1,5,10 seconds),
Battery Indicator (5 levels), Unit selection (Air Velocity: m/s, FPM Flow
Rate: m3/min, m3/h, CFM, cf/h Temperature: C, F Humidity: %RH
Static Pressure: kPa, Pa)
Store up to 25 Duct sizes (square/round) Maximum Data storage:
1500(equivalent to 277 hours)
Output: Digital Output*: RS-232C (Baud Rate 4800, 9600, 19200 & 39400 bps)
Analog Output*: DC 0-1V(select one from measuring parameters)
Power: 6 x AA Batteries
Battery Life: Approximately 10 hours. Continuous at 984fpm, 68°F with alkaline batteries
Operating Environment: Main: 41-104 °F
Probe: 32-140 °F
Storage Environment: 41-104 °F
Weight (main body): 0.9 lbs
Accessories: Carrying Case, Operation manual, AA batteries, Probe Cable