Professional Equipment     Print Page
PO Box 5197
Janesville, WI 53547-5197

Phone Number: 608-743-8002
Fax Number: 608-743-8039
Toll Free Number: 1-800-334-9291
Toll Free Fax Number: 1-888-776-3187


Fluke-707Ex Intrinsically Safe Loop Calibrator

Fluke-707Ex Intrinsically Safe Loop Calibrator
Item #: P609-7071
Manufacturer: Fluke Corporation
Manufacturer Item #: Fluke-707Ex
Product Type: Electrical Calibrators
Your Price: $875.00
Get quick & easy interpretation of readings with the Fluke-707Ex Intrinsically Safe Loop Calibrator
Features & Applications:

The 707Ex is an intrinsically safe loop calibrator for use in explosion endangered areas. The Fluke 707Ex is certified in accordance with the ATEX directive (Ex II 2 G Eex ia IIC T4) in Zones 1 and 2.

Features:

  • Large display and simple, quick-click push rotary button for easy one-handed operation
  • Simultaneous mA and % readout for quick, easy, interpretation of readings
  • mA accuracy of 0.015%, superior to other loop calibrators
  • 1 µA resolution for mA source, simulate and measure
  • Push button with 25% steps for fast, easy linearity checks
  • 0-100% "span check" for fast confirmation of zero and span
  • Selectable slow ramp, fast ramp, and step ramp to provide smooth outputs for valve slewing and loop functional tests
  • Internal loop supply, so you can power and read a transmitter at the same time without carrying a DMM
  • Measures Vdc to 28 V
  • Single 9 V battery that is easily changed
  • 0-20 mA or 4-20 mA default start up modes
  • Innovative output adjustment dial with 1 µA and 100 µA resolution
  • HART® compatible resistance is connected in series with the loop supply for compatibility with HART communicators

Spec Summary:

Specifications
Current Measurement:
Range: 0 to 24 mA
Resolution: 0.001 mA
Accuracy: 0.015% Rdg + 2 LSD
Current Sourcing:
Range: 0 to 20 mA or 4 to 20 mA
Accuracy: 0.015% Rdg + 2 LSD
Drive Capability: 700 Ù at 20 mA
Current Simulation:
Range: 0 to 20 mA or 4 to 20 mA
Accuracy: 0.015% Rdg + 2 LSD
Loop Power:
While measuring mA, 24 V
Voltage measurement:
Range: 0 to 28 V DC
Resolution: 1 mV
Accuracy: 0.015% Rdg + 2 LSD
Environmental Specifications
Operating Temperature:
14 to 131°F (-10 to 55°C)
Storage Temperature:
-40 to 140°F (-40 to 60°C)
Humidity (Without Condensation):
0 to 95% 32 to 86°F (0 to 30°C)
0 to 75% 86 to 104°F (30 to 40°C)
0 to 45% 104 to 122°F (40 to 50°C)
Temperature Coefficient:
-10 to 18°C, 28 to 55°C, ±.005% of range per °C
Safety Specifications
Safety:
CSA C22.2 No. 1010.1:1992
EMC:
EN50082-1:1992 and EN55022:1994 Class B
Maximum voltage:
30 V (protection circuit for 230 V)
ATEX certification:
Ex II G Eex ia IIC T4 Zones 1 and 2, ZELM 02 ATEX 0120 X
CE certification:
EN61010-1 and EN61326
Mechanical & General Specifications
Size:
5.98" x 2.75" x 1.61" (152 x 70 x 41 mm)
Weight:
7.9 oz (12.3 oz including holster)
[224 g (350 g including holster)]
Batteries:
One 9V battery Alkaline
Warranty:
One year
Shock & Vibration:
Per MIL-T-28800 for a Class 2 Instrument
Close Window